PANA 9600S Automatic Nucleic Acid Workstation
The PANA 9600S automatic nucleic acid workstation, which is designed and developed according to the principles of the magnetic beads method, is intended to greatly streamline the workflow for nucleic acid detection and extraction.
This workstation integrates the workflow of sample information scanning, sample loading, nucleic acid extraction, and PCR system set-up, etc. With compatible nucleic acid extraction kits, the required nucleic acids can be extracted from various common samples including whole blood, serum and plasma, swab and urine very quickly and efficiently for specific downstream applications.
-
-
01
One-key operation for modularized experiment flow: With one-key operation, automated sample information scanning, sample loading, nucleic acid extraction, and PCR system setup for 96 samples of novel coronavirus can be finished within 50min.allowing for rapid nucleic acid detection.
Minimized contamination measures: With rotary nucleic acid extraction, smart drop capture, strict zoning, air filtration, and UV disinfection technology, cross-contamination can be reduced to ensure accurate results.
High precision and reliable results:
With precise sample loading, accurate temperature ramp control, and precise liquid transfer design, consistent and precise results can be ensured for each of your assays. -
02
Highly flexible for your needs: Compatible with various sample types and extraction kits; 4 PCR systems can be set up at the same time.This flexibility allows it to meet diverse DNA and RNA extractions needs.
Automated workflow and hands-free operation: Automated barcode scanning for reagent identification, sample loading, nucleic acid extraction, PCR system setup: visualized consumable recognition; easy connection with LIS (laboratory information system).
Model | PANA9600S |
Sample Capacity | 1~96 |
Technical Principles | Magnetic beads method; Rotary nucleic acid extraction technology |
Processing Capacity | information scanning and nucleic acid extraction of 96 samples per run;4 different PCR system can be set up |
Sample Types | Plasma. serum, whole blood swab.and urine.etc. |
Sample Loading Channels | 4 |
Pipetting Performance | Below 15 uL: accuracy: A≤2.0%, repeatability: CV≤3.0%;15 uL to 50 uL: accuracy: A≤1.5%, repeatability: CV≤1.5%;Above 50 uL: accuracy: A≤1.0%, repeatability: CV≤1.0%. |
Liquid Level Detection | CapSense/Gas pressure sensor |
Sample Tubes | Compatible with all types of blood collection tube 1.5mL and 2.0mL centrifugal tubes, and sample loading cups, etc. |
Temperature Control | Lysis and elution, temperature flexible to control between 35°C and 120°C |
Information Tool | Barcode scanning for reagent identification; visualized consumable recognition |
PCR reagent chamber | Avoid light design; power-on automatic refrigeration (4°C ~15°C ) |
PCR consumables | Compatible with 0.1mL, 0.2mL 8 strip tube and 96-well plates |
Minimized Contamination | Independent closed extraction area,top directional exhaust creates an internal negative pressure system. Sampling device with air tightness and anti-dropping design External droplet catching plate Sterilizing device in experiment cabin and extraction cabin Customized function: directional ventilation system for the nucleic acid extraction area. |
Information Technology | Scanning the bar codes of multiple samples one by one while sample holder is loaded information connection of Sample tube-Deep well plate-PCR tube Easy connection with LIS (laboratory information system). |
Packaging Information | 1370mm(L)*810mm(W)*890mm(H):220kg(net);12-inch touch screen, muti-module real-time status monitor. |
Certificate | IVDR CE |
PANA9600S | |
---|---|
Sample Capacity | 1~96 |
Technical Principles | Magnetic beads method; Rotary nucleic acid extraction technology |
Processing Capacity | information scanning and nucleic acid extraction of 96 samples per run;4 different PCR system can be set up |
Sample Types | Plasma. serum, whole blood swab.and urine.etc. |
Sample Loading Channels | 4 |
Pipetting Performance | Below 15 uL: accuracy: A≤2.0%, repeatability: CV≤3.0%;15 uL to 50 uL: accuracy: A≤1.5%, repeatability: CV≤1.5%;Above 50 uL: accuracy: A≤1.0%, repeatability: CV≤1.0%. |
Liquid Level Detection | CapSense/Gas pressure sensor |
Sample Tubes | Compatible with all types of blood collection tube 1.5mL and 20mL centrifuaal tubes, freezing tubes, and sample loading cups, etc. |
Temperature Control | Lysis and elution, temperature flexible to control between 35°C and 120°C |
Information Tool | Barcode scanning for reagent identification; visualized consumable recognition |
PCR reagent chamber | Avoid light design; power-on automatic refrigeration (4°C ~15°C ) |
PCR consumables | Compatible with 0.1mL, 0.2mL 8 strip tube and 96-well plates |
Minimized Contamination | Independent closed extraction area,top directional exhaust creates an internal negative pressure system. Sampling device with air tightness and anti-dropping design External droplet catching plate Sterilizing device in experiment cabin and extraction cabin Customized function: directional ventilation system for the nucleic acid extraction area. |
Information Technology | Scanning the bar codes of multiple samples one by one while sample holder is loaded information connection of Sample tube-Deep well plate-PCR tube Easy connection with LIS (laboratory information system) |
Packaging Information | 1370mm(L)*810mm(W)*890mm(H):220kg(net);12-inch touch screen, muti-module real-time status monitor |
Certificate | IVDR CE |
Contact us